One year implanted patients follow up: SUAW
project first results
David
Guiraud1, Andrea Pacetti2, Elena Meola2, Jean
Louis Divoux3,
1 University of
2 Fondazione Montecatone
Onlus, Via Montecatone, 43 - 40026
3 MXM, 2720
4 Medicine Faculty, 2 Rue
Abstract
Two patients were implanted with a
sixteen channel neural/epimysial stimulator under the
European SUAW project. Restoring standing up and short distance walking in
paraplegics are the two main objectives. One year after the implantation, the
first patient is able to achieve both movements with acceptable performances.
During this year the system functioning and stimulation parameters were
monitored. The very first results obtained with no statistic relevance,
indicate tendencies and are close to the ones found in literature. The mix of
both types of stimulation is possible and suitable. It provides a new set of
useful data for the next generation of implantable
Introduction
Implanted
In the SUAW project, the aim was to
restore short distance walking and standing in complete paraplegics. Based on
literature and preliminary surface stimulation experiences, the surgical and
rehabilitation team chose a set of muscles stimulated by both epimysial and neural electrodes [2, 3, 7].
Materials
Implantable system
As we want to manage both epimysial and neural stimulation types, the system was
initially designed to be configurable in both by splitting the sixteen
available channels in two parts. This operation takes place at the very last
phase of integration on the PCB drawing. The system is therefore quite
flexible, and different implant configurations can be easily developed. For the
first two patients it was decided to provide four neural channels, and twelve epimysial ones with two references. This means that eight
wires for the neural, and fourteen wires for the epimysial
channels are needed. The characteristics of the SUAW implant are summarised in
table 1. Much more precise hardware characteristics are described in [1].
|
|
Typ. |
Max. |
Step |
|
|
Channels |
4
neural + 12 epimysial |
|||
|
I
neural |
2mA |
3.1mA
(3kW) |
50µA |
|
|
I epimysial |
10mA |
20mA
(1.5kW) 25.5mA
(1kW) |
100µA |
|
|
Pulse
width |
300µs |
1ms |
Continuous |
|
|
Frequency |
25Hz |
30
Hz |
Continuous |
|
Table 1: typical implant characteristics.
Figure 2: pulse waveform of neural and epimysial outputs: biphasic pulses with exponential
recovery phase.
The implant is also able to send back
pieces of information such as open circuit detection and internal power supply
level. The size is approximately 11x5x1.5cm. The electronic is packaged in a
ceramic box so that the HF antenna can be integrated in the implant itself
avoiding extra connections. The implant box is located above the umbilicus to
prevent it from migrating.
Control software and
external part
Power and data are transmitted
through HF link. The implant receives the stimulus parameters, intensity,
channel, and pulse width, and then
generates the corresponding pulse. A biprocessor
architecture is now available to develop the external programmer: a very
compact and flexible control unit about the size of a credit card. The power
consumption of this board is about 350mW. The HF module is currently redesigned
for enhanced performance, particularly regarding output transmission power
efficiency. The complete system synoptic is given on figure 3.

Figure 3: SUAW system synoptic
For the first trials, only open loop
movements are used, even if some closed loop control systems were used previous
to the implantation with surface stimulation. A PC software allows the
physiotherapist to program sequences of stimulation on each muscle with
trapezoidal amplitude shapes. Intensity or pulse width can be the variable
parameter and then the other is fixed at a predefined value for each muscle.
The stimulation frequency is common to the sixteen channels but will be
independent on the next software release. The biprocessor
board is able to keep all these patterns in flash memory, and then the system
becomes autonomous without any connection to a PC.
Methods
|
|
Patient A |
Patient B |
|
Age, country |
39, French |
30, Italian |
|
Sex |
Male |
Male |
|
Surgery |
28.09.1999 |
26.06.2000 |
|
Lesion |
T8 |
T5-T6 |
|
Muscles 1 2 3 4 5 6 7 |
Neural ch. foot drop quadriceps Epimysial ch gluteus maximus gluteus medius illiacus hamstrings none |
Neural ch. foot drop quadriceps Epimysial ch. gluteus maximus gluteus medius illiacus semi membranosis biceps |
|
Training |
3 times a week, 1 hour, from March to July
2000. 1 time a week, 2 hours, from September 2000 |
2 times a day from September 2000 |
Table 4: patient profile
The table 4 shows the patient
profiles decided by the medical team. Some epimysial
channels remain unused. The patients were regularly trained and the patient A
was able to stand up from nearly the beginning of training. As regards patient
A walking, the stimulation patterns are defined so as to achieve the following
sequence of movements : standing, swinging phase of one leg, standing. The
patient controls with push buttons the right or the left step. Physiotherapists
chose the intensity for the control parameter, and it is increased regularly
during the session for compensating fatigue.
Results
On the patient A technical problems
occurred on the implant itself but the overall design was not in question. In
March 2000 the patient was operated again in order to replace only the implant.
Since then, no technical problem has arisen on either patient.
Stimulation performances
Patient A is able to stand and walk,
and patient B is still under training. Two parameters are currently monitored:
the intensity thresholds, and the intensity levels needed to obtain the desired
movement. Force measurements were also done but are not presented in this
paper. A sample of the results are given in tables 5 and 6. Only results on one
neural channel (femoral nerve for quadriceps activation) and one epimysial channel (gluteus maximus)
are presented.
|
Dates |
26.06.2000, PW=250µs |
21.11.2000 PW=300µs |
|
Left quadriceps |
100µA |
450µA |
|
Right quadriceps |
100µA |
600µA |
|
Left gluteus |
2mA |
18mA |
|
Right gluteus |
3mA |
6mA |
Table 5: Patient B stimulation thresholds,
common stimulation frequency 20Hz.
|
Dates |
14.03.2000 |
18.04.2000 |
|
Left Q. |
2.55mA, 520µs |
2.6mA,600µs |
|
Right Q. |
2.9mA, 300µs |
2.4mA,600µs |
|
Left G. |
20mA,300µs |
25mA,300µs |
|
Right G. |
20mA,300µs |
25mA,500µs |
|
Standing/rest |
3x3min/5min |
15min/5min, 10min/5min |
Table 6: patient A standing parameters, values
at the end of the sessions, with 31Hz common stimulation frequency.
On both patients, the thresholds
increase with time, reflecting well known phenomena, among them fibrosis. The
same qualitative evolution can be observed on all the channels. On one hand,
the level of intensity needed for functioning at the beginning of the sessions
decreased with time. The training process can easily explain the improvement of
muscle performances. On the other hand, the values at the end of the sessions
have increased.
Concerning the movement achieved,
the patient B did not try to stand up, only training is in process. For the
patient A both standing and walking have been performed. Patient A is now able
to stand more than thirty minutes without too much fatigue. Patient A began
walking performance on April 2000. He is now able to walk in the lab on a four
meters long pathway for two hours, with five minutes of rest at each end of the
pathway. Actually, the patient goes two hours each week to the centre, and will
soon use the autonomous system. Sequence and muscle parameters during walking
will be presented during the session, nevertheless, the maximum values used are
close to the ones found during standing.
Discussion
From a qualitative point of view, we
have demonstrated that both stimulation types in the same implanted system are
possible. The opportunity of choosing neural or epimysial
stimulation types provides a great variety of configurations well adapted to
each patient. On one hand, neural channels give a high efficiency contraction
well suited for the quadriceps and an important enhancement of the walking
scheme with the foot drop stimulation. On the other hand, epimysial
stimulation is still needed on large muscles like glutei where the placement of
a neural electrode is very difficult. The ergonomic aspect of the system is
really improved compared to classical surface stimulation devices, and an
increase in performance is a partial consequence. The more frequently the
system is used, the more the muscle's reaction is improved. The next step
foreseen for 2001 is to provide to the patient a complete autonomous system to
use at home.
From a quantitative point of view,
results show that while thresholds increase, the performance of the muscles
does too. The result is that the control range is not large. The phenomenon is
higher on neural stimulation. Nevertheless neural stimulation provides a higher
stimulation efficiency but is less easy to control. In an open loop system a
fine control of the muscle's contraction is not really needed. Closed loop
control will give assessment about fine control through neural stimulation.
Fatigue resistance increases so that the duration of walking and standing
drastically changes. More than thirty minutes of standing and up to two hours
of walking with short periods of rest, provide sufficient autonomy to now let
the patient use the system at home.
The implant is versatile thanks to
its sixteen mixed channels but the surgery remains heavy. Not all the patients
are ready to accept this so that surface stimulation and orthotic
devices remain good solutions for some paraplegics. Implanted FES can greatly
be improved and the different experiences carried out through different teams
can not yet settle the best strategy to use. Waiting for the pure biological
solutions, implanted FES seems to be the best alternative in the near future.
References
[1] Guiraud
D., Denis B., Couderc P., Taroni
G., Stieglitz T., Description of a sixteen-channel
FES implantable system, IFESS2000, 18-21th of June, B55, Aalborg,
Denmark.
[2] M. Vincent,
M. Gaviria, G. Alba, F. Ohanna, Walking rehabilitation of the
paraplegic patient by means of surface electrical stimulation, IFESS2000.
[3] Alba G., Guiraud
D., The use of FES for gait rehabilitation in paraplegics, 5° conference of the
European Society for Engineering and Medicine, 6° international symposium of
Society of Biomedical Engineering and Telemedicine, Barcelona, 30 May-2 June
1999.
[4] Davis R., Eckouse
R., Patrick J. F., Delchanty A., Computerized
22-channel stimulator for limb movement, Applied Neurophysiology 1987.
[5] Donaldson N. D., Perkins T. A.,
Worley A. C. Lumbar root stimulation for restoring leg function: stimulator and
measurement of muscle actions, Artificial Organs, March 1997.
[6] Guiraud
D., Alba G., Gilbert B., From external to implanted FES, a high technology
product line, IV International conference on medical and biological new
technology, London, England, 1-3 September 1999.
[7] Kobetic
R., Triolo R. J., Marsolais
E. B., Muscle selection and walking performance of multichannel
FES systems for ambulation in paraplegia, IEEE Trans. on Rehabilitation
Engineering, 5, pp 23-29 1997.
[8] Kobetic
R; et al., Implanted functional electrical stimulation system for mobility in
paraplegia : a follow up case report, IEEE Trans. Rehabilitation Engineering,
December 1999.
Acknowledgements: this
research was carried out within the “Stand-Up-And-Walk” (SUAW) project. It has
been funded by a grant of the European Union within the BIOMED II program (BMH4-CT965-1501). We also
want to thank the first patients implanted, for their active cooperation. The
patients are cared for at the Centre Clémenceau, 45 Bd Clémenceau, F-67000
Strasbourg, and the Fondazione Montecatone.