Velocity-selective Recording using Multi-electrode Nerve Cuffs

 

Nick Donaldson

John Taylor and Jeff Winter

Dept. of Medical Physics & Bioengineering

University College London

11-20 Capper St, London, WC1E 6JA, UK

nickd@medphys.ucl.ac.uk

Dept. of Electronic Engineering

University College London

Torrington Place, London, WC1E 7JE, UK

j.taylor@ee.ucl.ac.uk

 

 


Abstract

A technique to improve the signal-to-noise ratio of recordings from cuff electrodes is presented. It is also shown that the system is inherently velocity selective. Unlike competing methods, the use of the system as a practical velocity-selective filter requires only relatively simple signal processing. This is due to the intrinsic properties of the multi-electrode cuff system.

 

Introduction

Hansen et al. [1] have shown that signals from cuff electrodes can be used chronically as inputs to neuroprostheses. This paper describes methods (a) for improving the signal-to-noise ratio (S/N) of neural signals picked up from cuffs and, (b) for discriminating between activities in different groups of neurons according to their propagation velocities.

 

Jezernik et al. [2] used signal processing of the signal from a tripolar cuff for these purposes: Wiener filters improved S/N; and the signal was classified according to activity in three types of receptor by autocorrelation and an artificial neural network. By contrast, our method employs a multi-electrode cuff to provide several tripolar signals, which are then subjected to relatively simple signal processing operations.

 

Improved Signal/Noise Ratio

Pflaum et al. [3] introduced an amplifier configuration for tripolar cuffs in which two front-end amplifiers are connected to the three electrodes (“true tripole”). That configuration is shown extended in Figure 1 for N+2 electrodes and N 2nd-rank amplifiers. If one action potential (AP) propagates through the cuff, it produces the same response at the outputs of all the 2nd-rank amplifiers, staggered in time by the electrode pitch divided by the velocity (T, 2T, 3T.. NT). Since T is velocity dependent, the artificial time delays (Nt...3t, 2t, t) will be matched to one propagation velocity. If they are matched, the AP responses are temporally re-aligned so that the output of the adder is greater than all the individual 2nd-rank amplifier responses by a factor N.

 

Assume that the length of the cuff is determined by anatomical and surgical considerations. The Riso & Pflaum true-tripole has three electrodes. As the number of electrodes is increased and hence the electrode pitch decreases, the signal amplitude at the outputs of each 2nd-rank amplifier decreases, but more signals are added together. Our theory predicts that there is a maximum in overall amplitude for a particular pitch corresponding to an optimal number of electrodes for a given cuff length [4].

 

Four types of noise are present: Johnson noise from the electrodes’ access resistances; Johnson noise from the axial resistances inside the cuff; amplifier voltage noise; and amplifier current noise. Analysis shows that when all these are appropriately summed, the total rises at a rate that is slightly less than ÖN.


 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 



For example, we considered a 30mm-long cuff. The maximal summed AP signal, for fibres of diameters 5, 10, 15 & 20 mm, occurs with 9 to 11 electrodes at a pitch of 3.75 to 3.00 mm. After the adder, the combined signal is 5 to 8 times larger, depending on fibre diameter, than from a single tripole (i.e. pitch 15 mm). For 11 electrodes, and assuming reasonable values for the resistances and amplifier noise characteristics, the predicted noise increases by a factor of 2.2 so the improvement in S/N is 2.3 to 3.6 (20 to 5 mm diameter, respectively).

 

Velocity Selectivity

Figure 1 shows functional blocks on several planes after the 2nd-rank amplifiers. Each plane has one output and time delays (Nt .…. 3t, 2t, t) that are matched to one propagation velocity, which may be positive or negative (afferent or efferent). On each plane there is also a narrow bandpass filter. A crucial part of this method is this: that the centre frequency of the passbands of these filters is proportionately related to the matched velocity. In a particular system with electrode pitch d and matched velocities (v1, v2, v3 ...vp), the artificial time delays for planes 1 to p will be t1=d/v1, t2=d/v2, t3=d/v3 ... tp=d/vp and the centre frequencies will be f1=v1/l, f2=v2/l, f3=v3/l....fp=vp/l, where l is a constant length which is be chosen by the designer. This proportionality means that the Q of the velocity-selective filter in the velocity domain is the same for all planes, which makes the filter outputs much easier to interpret. So if the delays (tp) in Fig 1 are chosen appropriately to cover the required velocity spectrum (probably in overlapping bands) and, in addition, the centre frequency of each


 

Figure 2: Outputs of three Velocity-Selective Filters

 


bandpass filter is chosen according to the above simple rule, then the relative activity in each velocity band can be measured directly.

 

Figure 2 shows a simulation result. Three velocity-domain filters have been chosen (3 planes), equally spaced in log(velocity), at 16.44, 34.64 and 73.00 m/s. APs propagate through an 11-electrode cuff every 0.02s. Their velocities, starting at t=0, are equally-spaced on a log scale from 10 to 120 m/s (10.0, 12.8, 16.4, 21.1, 27.0, 34.6, 44.4, 56.9, 73.0, 93.6 and 120). l = 50 m.

 

Summary

We expect that this method will allow velocity-selective recording and improved S/N in nerve cuff recordings. Activity in equal-sized fibres but propagating in opposite directions should be distinguishable, as should activity in different-sized fibres propagating in the same direction. To be practicable, we expect that the amplifiers must be mounted on the cuff, and this is an objective of the SENS Project (EU Grant QLG5-CT-2000-01372).

 

The invention described above is subject of a Patent Application (British Patent Office, Number 0201100.5, filed 18th January, 2002).

 

References

[1]  Hansen M., Haugland M., Sinkjaer T & Donaldson N. (2002) “Real time drop foot correction using machine learning and natural sensors.” Neuromodulation, 5, 41-53.

[2]  Jezernik S., Grill W. & Sinkjaer T. (1999) “Neurographic recordings, electrical stimulation and new neural signal processing methods for closed-loop neuroprosthetic control of bladder hyper-reflexia.” Proc. IFESS Conf., 81-84.

[3]  Pflaum C., Riso R. & Wiesspeiner (1995) “An improved nerve cuff recordings configuration for FES feedback control systems that utilise natural sensors.” Proc. IFESS Conf., 407-10.

[4]  Winter J., Rahal M., Taylor J., Donaldson N. de N. & Struijk J.J. (2000) “Improved spatial filtering of ENG signals using multielectrode nerve cuff.”Proc IFESS Conf., 368-71