Introduction
The use of accelerometers to measure
human motion has been suggested for over 30 years (Smidt
et al, 1971). With recent advances in miniaturization and the integration of
rate gyroscopes, portable motion sensors are becoming a viable alternative to
optical motion analysis systems (Mayagoitia et al,
2002). The development and refinement of accelerometer based motion sensors
were largely driven by the needs of
Neopraxis Ltd developed sensor packs that are part of the Neopraxis FES-22B stimulator system (Simcox et al, 2001). Individual sensors are extremely compact (12 x 34 x 65 mm), and light (25 g) (Simcox, et al, 2001). The sensors collect segmental orientation information (tilt, angular velocity and accelerations) in two orthogonal planes. Data are collected using a Pocket PC (Casio Cassiopeia EG800) fitted with a compact flash serial interface card. Multiple sensors can be attached to various limb segments and connected in series with the Pocket PC. This configuration allows for unprecedented portability and flexibility. The sensors also allow for attachment of external analogue devices (i.e. foots switches).
The purpose of this study was to
ascertain if the sensors, which were designed for
Methods
Gait Analysis: The methodology used to validate the sensors against a 3D
camera-based motion analysis system (CMAS) has been reported by Simcox et al (2001). However, while Simcox
et al, (2001) used 5 sensors, we used 7 and two pairs of foot switches (Force
Sensitive Resistors, MIE Ltd,
Protocol: At the beginning of testing, subjects were asked to stand as upright as possible with their back and heels touching the wall. Data were collected for 10 seconds. This was necessary to estimate sensor offset as the sensors, which are attached on the surface of each segment, are not parallel to the virtual segmental lines joining the proximal and distal landmarks. To obtain an estimate of the offset we assume that when a subject stands against the wall the thigh, shank and torso segments are orientated close to vertical (90°). The ankle joint is positioned neutral (0°). The offset readings are averaged and are subtracted in the gait analysis program. We acknowledge that our method to correct the offset may not be the most precise but we feel it is important to develop an easy and quick gait analysis system where an uncomplicated setup is an advantage.
Subjects were asked to walk at a
comfortable walking pace along a 10-meter walkway. The calibrated space within
the camera field of view is approximately 4 meters long, 1.5 meters wide and 2
meters high, enough for 2 to 3 strides. Between
Analysis: To analyze the gait data we developed special software in Labview (6i). We report data from the sagittal plane. Segmental tilt angles are converted into real world coordinates based on the individual’s anthropometric data and reconstructed into a biomechanical stick figure. Our program is based on frame-by-frame analysis and, combined with foot switch data, can analyze joint motion (ankle, knee, hip angles, angular velocity etc) and determine common spatiotemporal parameters (gait velocity, support phase, swing phase etc). Data from 3 right and 2 left strides were compared between CMAS and sensors. To compare data collected from different sampling frequencies both data sets were normalized to one gait cycle and spline interpolated to 100 points. The following data were analyzed: bilateral hip, knee and ankle joint angles, and gait velocity.
Balance Screening: The Neopraxis sensors are potentially ideal for the assessment of balance as they can directly measure sway. Clinical balance, or amount of sway, is assessed by measuring the center of pressure (COP) on the force plates. This approach assumes that the COP is representative of the total center of mass (COM) and represents sway. However, there are theoretical problems with this assumption. Sway by definition (Eng & Winter, 1993) is the movement of COM over the base of support and erroneous interpretations result if COP is used to measure sway (Eng & Winter, 1993). One sensor can measure sway directly if a sensor is placed on the dorsal aspect of torso close to the location of the COM (L2-3). The sensor measures angular deviations of the torso/COM in the sagittal and coronal planes. In this study, we measured sway by placing one sensor close to the location of the center of mass on the dorsal aspect of the torso (L2-3). Subject’s sway was assessed using a well-known clinical sensory organization test, called modified CTSIB.
Protocol: Ten normal subjects were tested under four conditions: standing on a firm surface with eyes open and then closed, standing on 5-inch thick foam with eyes open and then closed. One sensor was firmly attached on the dorsal surface over the vertebral column of the subject’s torso between levels L2-3. A small analog thumb trigger was connected to one connector port of the sensor. The subject was asked to stand still with one hand against the wall to increase stability. The experimenter stood behind the subject and started data collection. Sampling frequency was 57 Hz (18ms). After 5 seconds, the subject was asked to release their hand from the wall and start one of four conditions. Depending on the condition, specific instructions were given i.e. “step on the foam and when ready close your eyes and stand as still as possible until instructed to do otherwise.” Once the subject was ready, the trigger was depressed and 40 seconds of sway data were collected. The total test with setup lasted 3 to 5 minutes.
Analysis:
Data were downloaded onto a PC and analyzed using a
custom balance assessment program. The first 5 seconds of data (hand against
wall) prior to each balance condition were used to calculate sagittal and
coronal neutral angles. Sway data angles were subtracted from
the neutral angles so that all deviations were relative to 0º in
coronal and sagittal planes. Thirty seconds of sway data were analyzed after the trigger analog
signal. The coronal signal was filtered using a low-pass single order
Butterworth filter at 1 Hz. We found it not necessary to filter the sagittal
signal.
The following indices of sway were calculated from sagittal and coronal tilt angles: root mean error (RMS), average angular velocity (d/s) and the sum of all absolute angular deviations over 30s. Each measure is an indication of amount of sway. Larger numbers indicate more sway.
Results
Gait Analysis: Figure 1 shows the average joint angles and standard deviations for 5 strides (3 right and 2 left) obtained from the sensors and CMAS of 7 subjects. There is excellent agreement in terms of the shape of the waveforms and sub events. In fact, the sensor consistently picked up features in the gait cycle that were not picked up with the CMAS (i.e 1st rocker of the ankle joint). However, there still is considerable offset in the actual angular values. Some of the offset can be attributed to virtual segmental lines (joining the joint markers) not being entirely aligned and the choice use of the marker model. We will probably need to develop a more precise offset calculation protocol. However, the sensor data appear to agree more closely with standard gait reference data. The cross-correlations range between .92 and .99 for all subjects.
Gait velocity agreements were found to be excellent (1.45 ± .3 m/s sensors vs. 1.39 ± .2m/s CMAS), although the agreement depends very much on the accuracy of the anthropometric measures and marker placements. Event detection for CMAS was determined visually.
Balance Assessment: We report data for two sway scores: RMS and sway velocity (deg/s).
Two 2-factor analysis of variance (2 Eyes Open/Closed x 2 Support surface) show
that sway significantly increased as measured by RMS and sway velocity
increased when subjects stood on foam (RMS: F(1, 36)=23.842, p<.025; Sway
Velocity: F(1, 36) = 22.082, p<.025) as depicted in figure 2. There were no
other significant differences. There was no detectable sensor drift. The
results are consistent with what is reported in the literature. It suggests
that the sensors are sensitive enough to detect sway of small magnitudes. The
sensors are capable of resolving changes in tilt of .1º in both planes, which
compares favorably with CMAS calculated COM. This
demonstrates that the sensors are well suited to assess balance in a variety of
assessment tests (i.e. tandem walk, one-legged stance, limits of stability). In
addition, balance data can be collected quickly and at a fraction of the cost
of current force plate systems. In short, it is a perfect clinical tool.

Discussion / Conclusion:
The
results show that the sensors are well suited as a versatile and cost-effective
multipurpose motion analysis system; multiple sensors can be used for gait
analysis, while a single sensor can be used to assess balance. For each
application, the sensors report valid and reliable data. The sensors do not
require extensive laboratory facilities and can be used almost anywhere. At the
same time, we have also identified some limitations, for example, rigorous
activities such as running are beyond the current capability of the sensors
system hardware.
However,
the sensors form a solid design concept for a completely portable and
integrated motion analyzer. Specifically, the sensors
can also be integrated with other devices, such as heart rate monitors and foot
pressure distribution systems, allowing kinematic data to be correlated with
other measures (kinetic and physiological). Obviously, such a system, which
does not need dedicated laboratory facilities, will have many potential
applications in clinical, rehabilitation, sporting and research domains.
References:
[1] Eng, J. & Winter, D.A. (1993). Estimations of the horizontal displacement of the body centre of mass; considerations during standing activities. Gait & Posture, 1(3), 141-144.
[2] Simcox, S, Parker, S, Davis, G, Smith R, Middleton, JW. & Fisher, W. (2001). External limb orientation sensors for human movement. Proc. 5th Annual IFESS Conf. and Neural Prosthetics
[3] Smidt, GL, Arora, JS, & Johnston, RC.(1971). Accelegraphic analysis of several types of walking. Am J Phys Med, 50, 285-300.
[4] Mayagoitia, RE, Nene, AV, & Vetlink PH. (2002). Accelerometer and rate gyroscope measurement of kinematics: An inexpensive lternative to optical motion analysis systems. J Biomech, 35 537-542.
[5] Williamson, R, & Andrews, BJ. (2000). Sensor systems for lower limb functional electrical
stimulation (