On the development of modular miniaturized neural prostheses
Neural Prosthetics Group, Fraunhofer Institute for Biomedical Engineering, Ensheimer Strasse 48, D‑66386 St. Ingbert, Germany
Email: thomas.stieglitz@ibmt.fraunhofer.de
Website: www.ibmt.fraunhofer.de
Abstract
Neural prostheses have reached clinical practice to partially restore body functions and to support other therapies in neurological rehabilitation. However, new concepts often need decades to come from the first experimental results to clinical approval. Micromachining technologies paved the way to smaller and more complex neural implants but new materials often do not match the legal requirements for chronic implantations. Device approval with respect to national and international laws is time consuming and expensive. Therefore, we are joining a German initiative to develop modular miniaturized implants and standardized interfaces to speed up the development and decrease the costs. Within this paper, we describe our concept of interfaces and modules as well as the basics for a biocompatible assembling technology for neural microimplants.
1 Introduction
Neural prostheses in clinical practice may help patients to improve gait after hemiparesis, to manage the bladder function after spinal cord injury or to restore grasp in tetraplegia, for example [1]. The miniaturization of sensors and actuators and their integration with electronic circuits paved the way towards new applications in medicine for diagnosis, therapy, and rehabilitation within the last decades. Implantable biomedical microsystems need special assembling, encapsulation, and housing technologies to integrate mechanical, (bio-) electrical, and electronic components when titanium or ceramic housings are too space consuming and too stiff. International standards for physicochemical tests of implants with flexible substrates, encapsulation and housing materials that take care on the small amount of material and the “micro” dimensions of biomedical microimplants are neither approved nor established. The surface biocompatibility of materials was standardized several years ago [2] but aspects of structural biocompatibility with respect to the design, the weight and the flexibility of microimplants are still in the experimental state [3].
Taking the complexity of micromachining technology and the variety of anatomical and physiological needs and constraints into account, it might be a good idea to develop a “microsystem construction kit” with standardized modules and the description of interface standards. Within the “Initiative MicroMedicine” [4], we focused on demonstrators for telemetric micro systems with a special emphasis on neural prostheses in our group.
The
typical interfaces can be classified in four groups (Figure 1):
·
Interface I-0: functional interface between tissue and multiple sensors
for implantable systems. It includes the transducer performance as well as aspects
of biocompatible and biostable encapsulation.
·
Interface I-1: micro-opto-electro-mechanical
interface that connects multiple sensors with intelligent chip technology, e.g.
signal amplification and filtering.
·
Interface I-2: interface between chip technology and signal processing
circuitry and telemetric transceiver.
·
Interface I-3: wireless connection of an integrated implantable
telemetry unit with an external transmission and receiver station.
·
Interface I-4: external interface to a monitor or a telematic
data trasmission for telemonitoring,
telediagnostics or a virtual doctor’s round.

Figure
1: System
concept of a modular, telemetric in vivo sensor system or microimplant,
respectively, with core modules and interfaces I-0 to I-4 (modified from [4]).
2 Methods
The approach of a modular neural implant
was investigated on the example of a cuff electrode with an integrated
multiplexer to reduce the number of cables. The single modules were designed
and fabricated on different substrates. We chose polyimide as substrate and
insulation material with respect to its low weight, its flexibility and its
biocompatibility [5,6]. The polyimide PYRALIN PI 2611
(HD Microsystems,

Figure 2: Schematic view of a modular neural implant; electrodes and electronic circuitry on
flexible polyimide substrate.
A biocompatible and
stable assembling technology that is based on ball stud bonding, the so-called MicroFlex Interconnection- was used for chip assembly [8]
while gluing or soldering was used for SMD (capacitors, resistors). Cables were
fixed by welding on a screen printed alumina oxide ceramic as interface to the
thin-film substrate (Figure 2). Hybrid cuff electrodes were fabricated of
polyimide and silicone rubber and connected to the multiplexer via MicroFlex connections. The complete system was encapsulated
with silicone rubber. Functional tests were done in vitro to evaluate the
system concept and the biostability of the approach
[9].
3
Results
Micromachining
technologies were used to fabricate substrates and assemble the cuff electrodes
and multiplexer systems, respectively. The simple alignment of the connection
arrays of different modules under visual inspection led precisely set MicroFlex bonds between two thin and flexible substrates as
well as between chip and substrate (Figure 3).

Figure
3:
Biocompatible assembly of two modules; microsensor
(e.g. electrode) with signal processing module using double side “MicroFlex” ball stud technology.
Function tests proved
the electrical connection of the assemblies. After encapsulation in silicone
rubber, the 18 channel hybrid cuff electrode (6 tripoles)
with integrated multiplexer and Cooner wire cables
(Figure 4) worked in soaking tests in physiologic saline solution.

Figure
4: Modular
assembly of a cuff electrode
with multiplexer and Cooper cables.
4 Discussion and Conclusions
Flexible biomedical microimplants have been
fabricated for neural prostheses applications with a new modular approach.
First results have been promising. A modular assembly was feasible with a high
yield. With respect to a “microsystem construction
kit”, the interfaces must be specified in a more detailed way and general data
structures must be defined for an exchange between modules from different
research groups or distributors. Further on, the development and evaluation of
thin polymer-based encapsulations with respect to long-term biostability
is only at a very early development stage for biomedical microimplants.
Studies on the long-term stability and functionality of polyimide-based
biomedical microsystems for neural prostheses as
sieve-like interfaces to regenerating nerve stumps [10] and atraumatic
cuff-like interfaces to intact peripheral nerves [11] showed excellent
biocompatibility, biostability and functionality of
the implants in chronic animal models. The deposition of parylene
C on hybrid electronic assemblies improved the electrical insulation and biostability of the implants with no cytotoxic
side effects [12].
A modular approach to design and develop biomedical microimplants
for neural prostheses led to promising results with a first demonstrator. If
the standardization of interfaces and transmission protocols proceeds, this
approach might help to speed up the introduction of new neural prostheses into
clinical applications.
[1] Stieglitz T, Schüttler M., Koch KP, Neural Prostheses in Clinical Applications‑ Trends from Precision Mechanics towards Biomedical Microsystems in Neurological Rehabilitation, Biomed. Tech., 49: 72-77, 2004.
[2] ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing, 1997.
[3] Stieglitz, T, Considerations on Surface and Structural Biocompatibility as Prerequisite for Long-Term Stability of Neural Prostheses, J. Nanosci. Nanotechnol., 4: 1-8, 2004.
[4] url:-http://www.vde.com/mikromedizin (30.04.2004).
[5] Haggerty HS, Lusted HS, Histological Reaction to Polyimide Films in the Cochlea, Acta Otolaryngol., 107: 13-22, 1989.
[6]
[7] Stieglitz T, Beutel H, Schuettler M, et al., Micromachined, Polyimide-based Devices for Flexible Neural Interfaces, Biomed. Microdev., 2: 283-294, 2000.
[8]
Stieglitz T, Beutel H, Meyer JU, ‘Microflex‘‑A
New Assembling Technique for Interconnects, J. of Intel. Mat. Syst. Struct., 11: 417-426, 2000.
[9] Schuettler, M, Koch KP, Stieglitz T, et al., Multichannel Neural Cuff Electrodes with Integrated Multiplexer Circuit, Proceedings of the 1st Ann. Int. IEEE-EMBS Special Topic Conf. on Microtechnology in Medicine & Biology, 624-629, 2000.
[10] Ceballos D, Valero A, Valderrama E et al., Morphological and Functional Evaluation of Peripheral Nerve Fibers Regenerated through Polyimide Sieve Electrodes over Long-Term Implantation, J. Biomed. Mat. Res., 60: 517-528, 2002.
[11]
Rodríguez FJ, Ceballos D, Schuettler
M et al., Polyimide Cuff Electrodes for Peripheral Nerve Stimulation, J.
Neurosci. Met., 98: 105-118, 2000.
[12] Kammer S, Wien S, Koch KP et al., Investigations on the deposition of parylene C as encapsulation material for biomedical microimplants, Biomed. Techn. 47, Suppl. 1: 823-826, 2002.
Part of this work was funded by the German
Federal Ministry for Education and Research within the IMEX project (16SV1588). The authors thank Thomas Doerge and Sascha Kammer for their fabrication
of the implants and Dr. Martin Schuettler, (Implanted
Devices Group, University College London, UK) for many fruitful discussions
during the concept development and his work towards the first demonstrators.